  • ✇ICS Medical Advisories
  • NIHON KOHDEN Central Monitor CNS-6201 CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 8.7 ATTENTION: Exploitable remotely/low attack complexity Vendor: NIHON KOHDEN Equipment: Central Monitor CNS-6201 Vulnerability: NULL Pointer Dereference 2. RISK EVALUATION Successful exploitation of this vulnerability could allow an attacker to cause a denial-of-service condition. 3. TECHNICAL DETAILS 3.1 AFFECTED PRODUCTS The following NIHON KOHDEN products are affected: Central Monitor CNS-6201: Versions 01-03 Central Monitor CNS-6201: Versions 01-04
     

NIHON KOHDEN Central Monitor CNS-6201

Por:CISA
23 de Outubro de 2025, 11:31

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 8.7
  • ATTENTION: Exploitable remotely/low attack complexity
  • Vendor: NIHON KOHDEN
  • Equipment: Central Monitor CNS-6201
  • Vulnerability: NULL Pointer Dereference

2. RISK EVALUATION

Successful exploitation of this vulnerability could allow an attacker to cause a denial-of-service condition.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following NIHON KOHDEN products are affected:

  • Central Monitor CNS-6201: Versions 01-03
  • Central Monitor CNS-6201: Versions 01-04
  • Central Monitor CNS-6201: Versions 01-05
  • Central Monitor CNS-6201: Versions 01-06
  • Central Monitor CNS-6201: Versions 02-10
  • Central Monitor CNS-6201: Versions 02-11
  • Central Monitor CNS-6201: Versions 02-40

3.2 VULNERABILITY OVERVIEW

3.2.1 NULL POINTER DEREFERENCE CWE-476

In the affected versions of the CNS-6201 central monitor, if an attacker sends a specially crafted UDP packet, a NULL pointer dereference occurs during the receiving process, causing the process to terminate abnormally and resulting in a denial-of-service (DoS) condition. Exploiting this vulnerability does not require authentication, and it can be reproduced as long as the UDP service is reachable on the network.

CVE-2025-59668 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.5 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).

A CVSS v4 score has also been calculated for CVE-2025-59668. A base score of 8.7 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Japan

3.4 RESEARCHER

Jared P. Quinn of QuinnTech.ai reported the vulnerability to NIHON KOHDEN.

4. MITIGATIONS

The vulnerability identified in this advisory affects versions that are no longer supported. For the Central Monitor CNS-6201, maintenance support in the United States ended in September 2024 (EOS), and maintenance support for the relevant versions has already ended in other countries as well. NIHON KOHDEN recommends users migrate to successor products.

If users choose to continue using the CNS-6201 product, ensure the following supplementary controls are strictly implemented under the supervision of a network administrator:

  • Isolate the affected products from the Internet and any untrusted systems.
  • Monitor all network traffic attempting to reach the affected products and log any suspicious activity.
  • If integration with systems such as a Hospital Information System (HIS) is necessary, establish a boundary device (firewall or router) between the physically separated monitoring network and other networks, and explicitly allow only the minimum necessary communications.
  • Ensure monitoring redundancy by using a bedside monitor or medical telemetry system to establish an alternate network connection.

For more information, refer to the NIHON KOHDEN product security portal.

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:

  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time.

5. UPDATE HISTORY

  • October 23, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • Oxford Nanopore Technologies MinKNOW CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 8.3 ATTENTION: Exploitable remotely/low attack complexity Vendor: Oxford Nanopore Technologies Equipment: MinKNOW Vulnerabilities: Missing Authentication for Critical Function, Insufficiently Protected Credentials, Improper Check for Unusual or Exceptional Conditions 2. RISK EVALUATION Successful exploitation of these vulnerabilities could allow an attacker to disrupt sequencing operations and processes, exfiltrate and manipulate data, and bypass authenti
     

Oxford Nanopore Technologies MinKNOW

Por:CISA
21 de Outubro de 2025, 11:14

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 8.3
  • ATTENTION: Exploitable remotely/low attack complexity
  • Vendor: Oxford Nanopore Technologies
  • Equipment: MinKNOW
  • Vulnerabilities: Missing Authentication for Critical Function, Insufficiently Protected Credentials, Improper Check for Unusual or Exceptional Conditions

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker to disrupt sequencing operations and processes, exfiltrate and manipulate data, and bypass authentication controls.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following versions of MinKNOW, a DNA and RNA sequencing device, are affected:

  • MinKNOW: Versions prior to 24.06 (CVE-2024-35585)
  • MinKNOW: Versions prior to 24.11 (CVE-2025-54808, CVE-2025-10937)

3.2 VULNERABILITY OVERVIEW

3.2.1 MISSING AUTHENTICATION FOR CRITICAL FUNCTION CWE-306

A vulnerability exists in which remote access is enabled by default, and authentication relies on the IP address of the host computer. Unauthorized users on the same network can discover the IP address (e.g., via port scanning) and gain access to the sequencer by registering a legitimate or temporary Oxford Nanopore account. Once connected through the MinKNOW application, attackers can observe sequencing activity, pause or stop data collection, and redirect output data to an alternate location.

CVE-2024-35585 has been assigned to this vulnerability. A CVSS v3.1 base score of 8.6 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:L/A:L).

A CVSS v4 score has also been calculated for CVE-2024-35585. A base score of 8.3 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:L/VA:L/SC:N/SI:N/SA:N/R:U).

3.2.2 INSUFFICIENTLY PROTECTED CREDENTIALS CWE-522

Oxford Nanopore Technologies' MinKNOW software at or prior to version 24.11 stores authentication tokens in a file located in the system's temporary directory (/tmp) on the host machine. This directory is typically world-readable, allowing any local user or application to access the token. If the token is leaked (e.g., via malware infection or other local exploit), and remote access is enabled, it can be used to establish unauthorized remote connections to the sequencer. Remote access must be enabled for remote exploitation to succeed. This may occur either because the user has enabled remote access for legitimate operational reasons or because malware with elevated privileges (e.g., sudo access) enables it without user consent. This vulnerability can be chained with remote access capabilities to generate a developer token from a remote device. Developer tokens can be created with arbitrary expiration dates, enabling persistent access to the sequencer and bypassing standard authentication mechanisms.

CVE-2025-54808 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.8 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-54808. A base score of 7.3 has been calculated; the CVSS vector string is (AV:L/AC:L/AT:P/PR:L/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/R:U).

3.2.3 IMPROPER CHECK FOR UNUSUAL OR EXCEPTIONAL CONDITIONS CWE-754

Oxford Nanopore Technologies' MinKNOW software at or prior to version 24.11 creates a temporary file to store the local authentication token during startup, before copying it to its final location. This temporary file is created in a directory accessible to all users on the system. An unauthorized local user or process can exploit this behavior by placing a file lock on the temporary token file using the flock system call. This prevents MinKNOW from completing the token generation process. As a result, no valid local token is created, and the software is unable to execute commands on the sequencer. This leads to a denial-of-service (DoS) condition, blocking sequencing operations.

CVE-2025-10937 has been assigned to this vulnerability. A CVSS v3.1 base score of 5.5 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:L/UI:N/S:U/C:N/I:N/A:H).

A CVSS v4 score has also been calculated for CVE-2025-10937. A base score of 6.8 has been calculated; the CVSS vector string is (AV:L/AC:L/AT:N/PR:L/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/R:U).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: United Kingdom

3.4 RESEARCHER

Sara Rampazzi, Christina Boucher, Carson Stillman, Jonathan E. Bravo of the University of Florida reported these vulnerabilities to Oxford Nanopore Technologies.

4. MITIGATIONS

Oxford Nanopore Technologies recommends users upgrade to MinKNOW Versions later than 24.11 to eliminate these vulnerabilities.

If users are unable to upgrade to v24.11 to reduce risk from the remaining Authentication Token and Token Lock vulnerabilities, Oxford Nanopore advises the following additional measures for users on version 24.06:

  • Remote Connect: Keep Remote Connect disabled in MinKNOW unless strictly required, and enable it only within trusted network environments.
  • Endpoint Protection: Install and maintain antivirus and malware scanning tools to mitigate denial-of-service (DoS) conditions arising from local exploitation or malware.

Users running older versions of MinKNOW who cannot upgrade immediately should contact Oxford Nanopore Support for guidance on securing their configurations. Downloading the release requires users to be logged into the Nanopore Community.

CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time.

5. UPDATE HISTORY

  • October 21, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • FUJIFILM Healthcare Americas Synapse Mobility CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 5.3 ATTENTION: Exploitable remotely/low attack complexity Vendor: FUJIFILM Healthcare Americas Corporation Equipment: Synapse Mobility Vulnerability: External Control of Assumed-Immutable Web Parameter 2. RISK EVALUATION Successful exploitation of this vulnerability could allow an attacker to access information beyond their assigned roles. 3. TECHNICAL DETAILS 3.1 AFFECTED PRODUCTS The following versions of FUJIFILM Healthcare Americas Synapse Mobility ar
     

FUJIFILM Healthcare Americas Synapse Mobility

Por:CISA
21 de Agosto de 2025, 09:06

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 5.3
  • ATTENTION: Exploitable remotely/low attack complexity
  • Vendor: FUJIFILM Healthcare Americas Corporation
  • Equipment: Synapse Mobility
  • Vulnerability: External Control of Assumed-Immutable Web Parameter

2. RISK EVALUATION

Successful exploitation of this vulnerability could allow an attacker to access information beyond their assigned roles.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following versions of FUJIFILM Healthcare Americas Synapse Mobility are affected:

  • Synapse Mobility: Versions prior to 8.2

3.2 VULNERABILITY OVERVIEW

3.2.1 EXTERNAL CONTROL OF ASSUMED-IMMUTABLE WEB PARAMETER CWE-472

FUJIFILM Healthcare Americas Synapse Mobility versions prior to 8.2 contains a privilege escalation vulnerability through external control of Web parameter. Exploitation of this vulnerability could allow an attacker to bypass authentication and access information beyond role-based access controls.

CVE-2025-54551 has been assigned to this vulnerability. A CVSS v3.1 base score of 4.3 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:N/A:N).

A CVSS v4 score has also been calculated for CVE-2025-54551. A base score of 5.3 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:N/PR:L/UI:N/VC:L/VI:N/VA:N/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Japan

3.4 RESEARCHER

Christopher Alejandro (Moroco) reported this vulnerability to CISA.

4. MITIGATIONS

FUJIFILM Healthcare Americas has released the following mitigation options for users to follow:

  • It is recommended to upgrade to the latest version of the product or any version from 8.2 and later to resolve this issue.
  • There is an immediate mitigation available of disabling the search function in the configurator settings. Access to the search function for all users can also be removed by unchecking the "Allow plain text accession number" check box in the security section of the admin interface. This will limit the site to use of the product only via the SecureURL feature.
  • Patches have been released which will remediate the vulnerability for software versions 8.0-8.1.1.

FUJIFILM Healthcare Americas strongly encourages users to update the Synapse Mobility application to the latest version if they are past their end of support date.

For more information, refer to Synapse Mobility Vulnerability Notification.

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time.

5. UPDATE HISTORY

  • August 21, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • Santesoft Sante PACS Server CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 9.1 ATTENTION: Exploitable remotely/low attack complexity Vendor: Santesoft Equipment: Sante PACS Server Vulnerabilities: Path Traversal, Double Free, Cleartext Transmission of Sensitive Information, Cross-site Scripting 2. RISK EVALUATION Successful exploitation of these vulnerabilities could allow an attacker to create arbitrary files, cause a denial-of-service condition, obtain sensitive information, and steal a user's cookie information. 3. TECHNICAL
     

Santesoft Sante PACS Server

Por:CISA
12 de Agosto de 2025, 09:54

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 9.1
  • ATTENTION: Exploitable remotely/low attack complexity
  • Vendor: Santesoft
  • Equipment: Sante PACS Server
  • Vulnerabilities: Path Traversal, Double Free, Cleartext Transmission of Sensitive Information, Cross-site Scripting

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker to create arbitrary files, cause a denial-of-service condition, obtain sensitive information, and steal a user's cookie information.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Santesoft products are affected:

  • Sante PACS Server: Versions prior to 4.2.3

3.2 VULNERABILITY OVERVIEW

3.2.1 IMPROPER LIMITATION OF A PATHNAME TO A RESTRICTED DIRECTORY ('PATH TRAVERSAL') CWE-22

Sante PACS Server Web Portal allows remote attackers utilize DCM files to create arbitrary files on affected installations of Sante PACS Server.

CVE-2025-0572 has been assigned to this vulnerability. A CVSS v3.1 base score of 4.3 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:L/UI:N/S:U/C:N/I:L/A:N).

A CVSS v4 score has also been calculated for CVE-2025-0572. A base score of 5.3 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:N/PR:L/UI:N/VC:N/VI:L/VA:N/SC:N/SI:N/SA:N).

3.2.2 DOUBLE FREE CWE-415

The Sante PACS Server allows a remote attacker to crash the main thread by sending a crafted HL7 message, causing a denial-of-service condition. The application would require a manual restart and no authentication is required.

CVE-2025-53948 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.5 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).

A CVSS v4 score has also been calculated for CVE-2025-53948. A base score of 8.7 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N).

3.2.3 CLEARTEXT TRANSMISSION OF SENSITIVE INFORMATION CWE-319

The Sante PACS Server Web Portal sends credential information without encryption.

CVE-2025-54156 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.4 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:N).

A CVSS v4 score has also been calculated for CVE-2025-54156. A base score of 9.1 has been calculated; the CVSS vector string is (AV:N/AC:H/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N).

3.2.4 IMPROPER NEUTRALIZATION OF INPUT DURING WEB PAGE GENERATION ('CROSS-SITE SCRIPTING') CWE-79

Sante PACS Server web portal is vulnerable to stored cross-site scripting. An attacker could inject malicious HTML codes redirecting a user to a malicious webpage and stealing the user's cookie.

CVE-2025-54862 has been assigned to this vulnerability. A CVSS v3.1 base score of 5.4 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:L/UI:R/S:C/C:L/I:L/A:N).

A CVSS v4 score has also been calculated for CVE-2025-54862. A base score of 4.8 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:N/PR:L/UI:A/VC:L/VI:L/VA:N/SC:L/SI:L/SA:N).

3.2.5 IMPROPER NEUTRALIZATION OF INPUT DURING WEB PAGE GENERATION ('CROSS-SITE SCRIPTING') CWE-79

Sante PACS Server is vulnerable to stored cross-site scripting. An attacker could inject malicious HTML codes redirecting a user to a malicious webpage and stealing the user's cookie.

CVE-2025-54759 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.1 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:C/C:L/I:L/A:N).

A CVSS v4 score has also been calculated for CVE-2025-54759. A base score of 5.1 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:N/PR:N/UI:A/VC:L/VI:L/VA:N/SC:L/SI:L/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Cyprus

3.4 RESEARCHER

Chizuru Toyama of TXOne Networks reported these vulnerabilities to CISA.

4. MITIGATIONS

Santesoft recommends users update PACS Server to Version 4.2.3 or later.

CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the Internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time.

5. UPDATE HISTORY

  • August 12, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • Medtronic MyCareLink Patient Monitor CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 7.0 ATTENTION: Low attack complexity Vendor: Medtronic Equipment: MyCareLink Patient Monitor 24950, 24952 Vulnerabilities: Cleartext Storage of Sensitive Information, Empty Password in Configuration File, Deserialization of Untrusted Data 2. RISK EVALUATION Successful exploitation of these vulnerabilities could lead to system compromise, unauthorized access to sensitive data, and manipulation of the monitor's functionality. 3. TECHNICAL DETAILS 3.1 AFFECT
     

Medtronic MyCareLink Patient Monitor

Por:CISA
24 de Julho de 2025, 11:26

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 7.0
  • ATTENTION: Low attack complexity
  • Vendor: Medtronic
  • Equipment: MyCareLink Patient Monitor 24950, 24952
  • Vulnerabilities: Cleartext Storage of Sensitive Information, Empty Password in Configuration File, Deserialization of Untrusted Data

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could lead to system compromise, unauthorized access to sensitive data, and manipulation of the monitor's functionality.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Medtronic products are affected:

  • MyCareLink Patient Monitor model 24950: All versions
  • MyCareLink Patient Monitor model 24952: All versions

3.2 Vulnerability Overview

3.2.1 CLEARTEXT STORAGE OF SENSITIVE INFORMATION CWE-312

Medtronic MyCareLink Patient Monitor uses an unencrypted filesystem on internal storage, which allows an attacker with physical access to read and modify files.

CVE-2025-4394 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.8 has been calculated; the CVSS vector string is (AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-4394. A base score of 7.0 has been calculated; the CVSS vector string is (AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.2.2 EMPTY PASSWORD IN CONFIGURATION FILE CWE-258

Medtronic MyCareLink Patient Monitor has a built-in user account with an empty password, which allows an attacker with physical access to log in with no password and access/modify system functionality.

CVE-2025-4395 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.8 has been calculated; the CVSS vector string is (AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-4395. A base score of 7.0 has been calculated; the CVSS vector string is (AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.2.3 DESERIALIZATION OF UNTRUSTED DATA CWE-502

Medtronic MyCareLink Patient Monitor has an internal service that deserializes data, which allows a local attacker to interact with the service by crafting a binary payload to crash the service or elevate privileges.

CVE-2025-4393 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.5 has been calculated; the CVSS vector string is (AV:L/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-4393. A base score of 5.9 has been calculated; the CVSS vector string is (AV:L/AC:H/AT:P/PR:L/UI:N/VC:L/VI:H/VA:H/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Ireland

3.4 RESEARCHER

Ethan Morchy from Somerset Recon and Carl Mann, an independent researcher, reported these vulnerabilities to Medtronic.

4. MITIGATIONS

The identified vulnerabilities were reported as low-risk findings. An attacker would need to physically tamper with the monitor to exploit them. In response, starting in June 2025, Medtronic began deploying security updates to address these findings.

Medtronic recommends the following actions:

  • The security update process is performed automatically when the monitor is connected to the internet. Users should ensure that their remote monitor is plugged in to receive updates.
  • Physicians should continue to prescribe monitors as intended.
  • Users should maintain possession of their home monitor.
  • Users should only use home monitors provided directly from a healthcare provider or a Medtronic representative.

Users needing additional assistance should contact security@medtronic.com.

For more information regarding these vulnerabilities, refer to Medtronic's security bulletin.

Users should follow CISA's guidance in the following areas:

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time. These vulnerabilities are not exploitable remotely.

5. UPDATE HISTORY

  • July 24, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • Panoramic Corporation Digital Imaging Software CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 8.5 ATTENTION: Low attack complexity Vendor: Panoramic Corporation Equipment: Digital Imaging Software Vulnerability: Uncontrolled Search Path Element 2. RISK EVALUATION Successful exploitation of this vulnerability could allow a standard user to obtain NT Authority/SYSTEM privileges. 3. TECHNICAL DETAILS 3.1 AFFECTED PRODUCTS The following Panoramic Corporation products are affected: Digital Imaging Software: Version 9.1.2.7600 3.2 VULNERABILITY OVERVI
     

Panoramic Corporation Digital Imaging Software

Por:CISA
17 de Julho de 2025, 09:20

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 8.5
  • ATTENTION: Low attack complexity
  • Vendor: Panoramic Corporation
  • Equipment: Digital Imaging Software
  • Vulnerability: Uncontrolled Search Path Element

2. RISK EVALUATION

Successful exploitation of this vulnerability could allow a standard user to obtain NT Authority/SYSTEM privileges.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Panoramic Corporation products are affected:

  • Digital Imaging Software: Version 9.1.2.7600

3.2 VULNERABILITY OVERVIEW

3.2.1 UNCONTROLLED SEARCH PATH ELEMENT CWE-427

The affected product is vulnerable to DLL hijacking, which may allow an attacker to obtain NT Authority/SYSTEM as a standard user.

CVE-2024-22774 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.8 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2024-22774. A base score of 8.5 has been calculated; the CVSS vector string is (AV:L/AC:L/AT:N/PR:L/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: North America
  • COMPANY HEADQUARTERS LOCATION: United States

3.4 RESEARCHER

Damian Semon Jr. of Blue Team Alpha LLC reported this vulnerability to CISA.

4. MITIGATIONS

The affected software is vulnerable due to an SDK component owned by Oy Ajat Ltd, which is no longer supported. Panoramic Corporation is not the owner of this vulnerable component. Panoramic Corporation did not recommend any specific mitigation for this vulnerability. Users should contact Panoramic Corporation's support address for further information.

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the Internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time. This vulnerability is not exploitable remotely.

5. UPDATE HISTORY

  • July 17, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • MicroDicom DICOM Viewer CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 8.6 ATTENTION: Exploitable remotely/low attack complexity Vendor: MicroDicom Equipment: DICOM Viewer Vulnerability: Out-of-bounds Write 2. RISK EVALUATION Successful exploitation of this vulnerability could allow an attacker to execute arbitrary code on affected installations of DICOM Viewer. 3. TECHNICAL DETAILS 3.1 AFFECTED PRODUCTS The following MicroDicom products are affected: DICOM Viewer: Versions 2025.2 (Build 8154) and prior 3.2 VULNERABILITY O
     

MicroDicom DICOM Viewer

Por:CISA
10 de Junho de 2025, 07:13

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 8.6
  • ATTENTION: Exploitable remotely/low attack complexity
  • Vendor: MicroDicom
  • Equipment: DICOM Viewer
  • Vulnerability: Out-of-bounds Write

2. RISK EVALUATION

Successful exploitation of this vulnerability could allow an attacker to execute arbitrary code on affected installations of DICOM Viewer.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following MicroDicom products are affected:

  • DICOM Viewer: Versions 2025.2 (Build 8154) and prior

3.2 VULNERABILITY OVERVIEW

3.2.1 OUT-OF-BOUNDS WRITE CWE-787

DICOM Viewer suffers from an out-of-bounds write vulnerability. Remote attackers are able to exploit this issue to potentially execute arbitrary code on affected installations of DICOM Viewer. User interaction is required to exploit the vulnerability in that the user must either visit a malicious website or open a malicious DICOM file locally.

CVE-2025-5943 has been assigned to this vulnerability. A CVSS v3.1 base score of 8.8 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-5943. A base score of 8.6 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:N/PR:N/UI:A/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Bulgaria

3.4 RESEARCHER

Michael Heinzl reported this vulnerability to CISA.

4. MITIGATIONS

MicroDicom recommends users update DICOM Viewer to version 2025.3 or later.

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time.

5. UPDATE HISTORY

  • June 10, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • Santesoft Sante DICOM Viewer Pro CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 8.4 ATTENTION: Low attack complexity Vendor: Santesoft Equipment: Sante DICOM Viewer Pro Vulnerability: Out-of-bounds Read 2. RISK EVALUATION Successful exploitation of this vulnerability could allow an attacker to disclose information or execute arbitrary code. 3. TECHNICAL DETAILS 3.1 AFFECTED PRODUCTS The following Santesoft products are affected: Sante DICOM Viewer Pro: Versions 14.2.1 and prior 3.2 VULNERABILITY OVERVIEW 3.2.1 OUT-OF-BOUNDS READ CW
     

Santesoft Sante DICOM Viewer Pro

Por:CISA
29 de Maio de 2025, 10:02

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 8.4
  • ATTENTION: Low attack complexity
  • Vendor: Santesoft
  • Equipment: Sante DICOM Viewer Pro
  • Vulnerability: Out-of-bounds Read

2. RISK EVALUATION

Successful exploitation of this vulnerability could allow an attacker to disclose information or execute arbitrary code.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Santesoft products are affected:

  • Sante DICOM Viewer Pro: Versions 14.2.1 and prior

3.2 VULNERABILITY OVERVIEW

3.2.1 OUT-OF-BOUNDS READ CWE-125

The affected product contains a memory corruption vulnerability. A local attacker could exploit this issue to potentially disclose information and to execute arbitrary code on affected installations of Sante DICOM Viewer Pro.

CVE-2025-5307 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.8 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-5307. A base score of 8.4 has been calculated; the CVSS vector string is (AV:L/AC:L/AT:N/PR:N/UI:A/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Cyprus

3.4 RESEARCHER

Michael Heinzl reported this vulnerability to CISA.

4. MITIGATIONS

Santesoft recommends users upgrade Sante DICOM Viewer Pro to version v14.2.2.

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the Internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time. This vulnerability is not exploitable remotely.

5. UPDATE HISTORY

  • May 29, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • Pixmeo OsiriX MD CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 9.3 ATTENTION: Exploitable remotely/low attack complexity Vendor: Pixmeo Equipment: OsiriX MD Vulnerabilities: Use After Free, Cleartext Transmission of Sensitive Information 2. RISK EVALUATION Successful exploitation of these vulnerabilities could allow an attacker to cause memory corruption, resulting in a denial-of-service condition or to steal credentials. 3. TECHNICAL DETAILS 3.1 AFFECTED PRODUCTS The following Pixmeo products are affected: OsiriX M
     

Pixmeo OsiriX MD

Por:CISA
8 de Maio de 2025, 08:56

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 9.3
  • ATTENTION: Exploitable remotely/low attack complexity
  • Vendor: Pixmeo
  • Equipment: OsiriX MD
  • Vulnerabilities: Use After Free, Cleartext Transmission of Sensitive Information

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker to cause memory corruption, resulting in a denial-of-service condition or to steal credentials.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Pixmeo products are affected:

  • OsiriX MD: Versions 14.0.1 (Build 2024-02-28) and prior

3.2 VULNERABILITY OVERVIEW

3.2.1 USE AFTER FREE CWE-416

The affected product is vulnerable to a use after free scenario, which could allow an attacker to upload a crafted DICOM file and cause memory corruption leading to a denial-of-service condition.

CVE-2025-27578 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.5 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).

A CVSS v4 score has also been calculated for CVE-2025-27578. A base score of 8.7 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N).

3.2.2 USE AFTER FREE CWE-416

The affected product is vulnerable to a local use after free scenario, which could allow an attacker to locally import a crafted DICOM file and cause memory corruption or a system crash.

CVE-2025-31946 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.2 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).

A CVSS v4 score has also been calculated for CVE-2025-31946. A base score of 6.9 has been calculated; the CVSS vector string is (AV:L/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N).

3.2.3 CLEARTEXT TRANSMISSION OF SENSITIVE INFORMATION CWE-319

The Osirix MD Web Portal sends credential information without encryption, which could allow an attacker to steal credentials.

CVE-2025-27720 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.4 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:N).

A CVSS v4 score has also been calculated for CVE-2025-27720. A base score of 9.3 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Switzerland

3.4 RESEARCHER

Chizuru Toyama of TXOne Networks and Canaan Kao of TXOne Networks reported these vulnerabilities to CISA.

4. MITIGATIONS

Pixmeo recommends users to download the latest version of OsiriX MD.

For additional support regarding OsiriX MD, users should contact Pixmeo directly.

CISA recommends users take defensive measures to minimize the risk of exploitation of this these vulnerabilities, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the Internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time.

5. UPDATE HISTORY

  • May 8, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • MicroDicom DICOM Viewer CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 8.6 ATTENTION: Exploitable remotely/low attack complexity Vendor: MicroDicom Equipment: DICOM Viewer Vulnerabilities: Out-of-Bounds Write, Out-of-Bounds Read 2. RISK EVALUATION Successful exploitation of these vulnerabilities could allow an attacker to disclose information, cause memory corruption, and execute arbitrary code. 3. TECHNICAL DETAILS 3.1 AFFECTED PRODUCTS The following MicroDicom products are affected: DICOM Viewer: Versions 2025.1 (Build 33
     

MicroDicom DICOM Viewer

Por:CISA
1 de Maio de 2025, 09:17

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 8.6
  • ATTENTION: Exploitable remotely/low attack complexity
  • Vendor: MicroDicom
  • Equipment: DICOM Viewer
  • Vulnerabilities: Out-of-Bounds Write, Out-of-Bounds Read

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker to disclose information, cause memory corruption, and execute arbitrary code.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following MicroDicom products are affected:

  • DICOM Viewer: Versions 2025.1 (Build 3321) and prior

3.2 VULNERABILITY OVERVIEW

3.2.1 OUT-OF-BOUNDS WRITE CWE-787

MicroDicom DICOM Viewer is vulnerable to an out-of-bounds write which may allow an attacker to execute arbitrary code. The user must open a malicious DCM file for exploitation.

CVE-2025-35975 has been assigned to this vulnerability. A CVSS v3.1 base score of 8.8 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-35975. A base score of 8.6 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:N/PR:N/UI:A/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.2.2 OUT-OF-BOUNDS READ CWE-125

MicroDicom DICOM Viewer is vulnerable to an out-of-bounds read which may allow an attacker to cause memory corruption within the application. The user must open a malicious DCM file for exploitation.

CVE-2025-36521 has been assigned to this vulnerability. A CVSS v3.1 base score of 8.8 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-36521. A base score of 8.6 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:N/PR:N/UI:A/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Bulgaria

3.4 RESEARCHER

Michael Heinzl reported these vulnerabilities to CISA.

4. MITIGATIONS

MicroDicom recommends user update DICOM Viewer to version 2025.2 or later.

CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time.

5. UPDATE HISTORY

  • May 1, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • INFINITT Healthcare INFINITT PACS CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 8.7 ATTENTION: Exploitable remotely/low attack complexity/public exploits are available Vendor: INFINITT Healthcare Equipment: INFINITT PACS Vulnerabilities: Unrestricted Upload of File with Dangerous Type, Exposure of Sensitive System Information to an Unauthorized Control Sphere, 2. RISK EVALUATION Successful exploitation of these vulnerabilities could allow an attacker to upload malicious files and access unauthorized system resources, resulting in arb
     

INFINITT Healthcare INFINITT PACS

Por:CISA
10 de Abril de 2025, 12:19

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 8.7
  • ATTENTION: Exploitable remotely/low attack complexity/public exploits are available
  • Vendor: INFINITT Healthcare
  • Equipment: INFINITT PACS
  • Vulnerabilities: Unrestricted Upload of File with Dangerous Type, Exposure of Sensitive System Information to an Unauthorized Control Sphere,

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker to upload malicious files and access unauthorized system resources, resulting in arbitrary code execution or information disclosure.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following INFINITT Healthcare products are affected:

  • INFINITT PACS System Manager: Versions 3.0.11.5 BN9 and prior

3.2 VULNERABILITY OVERVIEW

3.2.1 UNRESTRICTED UPLOAD OF FILE WITH DANGEROUS TYPE CWE-434

An attacker could exploit this vulnerability by uploading arbitrary files via the a specific endpoint, leading to unauthorized remote code execution or system compromise.

CVE-2025-27714 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.3 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:L).

A CVSS v4 score has also been calculated for CVE-2025-27714. A base score of 5.3 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:N/PR:L/UI:N/VC:L/VI:L/VA:L/SC:N/SI:N/SA:N).

3.2.2 UNRESTRICTED UPLOAD OF FILE WITH DANGEROUS TYPE CWE-434

An attacker could exploit this vulnerability by uploading arbitrary files via a specific service, which could lead to system compromise.

CVE-2025-24489 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.3 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:L).

A CVSS v4 score has also been calculated for CVE-2025-24489. A base score of 5.3 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:N/PR:L/UI:N/VC:L/VI:L/VA:L/SC:N/SI:N/SA:N).

3.2.3 EXPOSURE OF SENSITIVE SYSTEM INFORMATION TO AN UNAUTHORIZED CONTROL SPHERE CWE-497

Unauthorized users can access the system without proper authorization, which could lead to unauthorized access to system resources.

CVE-2025-27721 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.5 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N).

A CVSS v4 score has also been calculated for CVE-2025-27721. A base score of 8.7 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: South Korea

3.4 RESEARCHER

Piotr Kijewski of the Shadowserver Foundation reported these vulnerabilities to CISA.

4. MITIGATIONS

INFINITT recommends the following mitigations:

The latest version of the software (3.0.11.5 BN10 or later) is NOT affected, as it includes default security patches.

INFINITT ULite is NOT affected by these vulnerabilities. However, if INFINITT ULite is operating as an integrated system with INFINITT PACS, patching is required to secure the PACS environment.

  • For CVE-2025-27714 and CVE-2025-24489: Apply the security patch and configure the System Manager settings to restrict unauthorized file uploads.
  • For CVE-2025-27721: Apply the patch, enforce strong password policies, and enable logging to monitor for unauthorized access attempts.
  • Network Security Recommendations: Minimize network exposure for PACS servers, ensuring they are not directly accessible from the internet.
  • Contact Information: Customers requiring additional support should contact INFINITT Security Team. (cybersecurity@infinitt.com)

CISA recommends users take defensive measures to minimize the risk of exploitation of this these vulnerabilities, such as:

  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

5. UPDATE HISTORY

  • April 10, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • Santesoft Sante DICOM Viewer Pro CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 8.4 ATTENTION: Low attack complexity Vendor: Santesoft Equipment: Sante DICOM Viewer Pro Vulnerability: Out-of-Bounds Write 2. RISK EVALUATION Successful exploitation of this vulnerability could allow an attacker to cause memory corruption that would result in execution of arbitrary code. 3. TECHNICAL DETAILS 3.1 AFFECTED PRODUCTS The following Santesoft products are affected: Sante DICOM Viewer Pro: Versions 14.1.2 and prior 3.2 VULNERABILITY OVERVIEW
     

Santesoft Sante DICOM Viewer Pro

Por:CISA
20 de Março de 2025, 08:55

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 8.4
  • ATTENTION: Low attack complexity
  • Vendor: Santesoft
  • Equipment: Sante DICOM Viewer Pro
  • Vulnerability: Out-of-Bounds Write

2. RISK EVALUATION

Successful exploitation of this vulnerability could allow an attacker to cause memory corruption that would result in execution of arbitrary code.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Santesoft products are affected:

  • Sante DICOM Viewer Pro: Versions 14.1.2 and prior

3.2 VULNERABILITY OVERVIEW

3.2.1 OUT-OF-BOUNDS WRITE CWE-787

The affected product is vulnerable to an out-of-bounds write, which requires a user to open a malicious DCM file, resulting in execution of arbitrary code by a local attacker.

CVE-2025-2480 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.8 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-2480. A base score of 8.4 has been calculated; the CVSS vector string is (AV:L/AC:L/AT:N/PR:N/UI:A/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Cyprus

3.4 RESEARCHER

Michael Heinzl reported this vulnerability to CISA.

4. MITIGATIONS

Santesoft released an updated version of their product and recommends updating Sante DICOM Viewer Pro to v14.2.0 or later.

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time. This vulnerability is not exploitable remotely.

5. UPDATE HISTORY

  • March 20, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • Philips Intellispace Cardiovascular (ISCV) CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 8.5 ATTENTION: Low attack complexity Vendor: Philips Equipment: Intellispace Cardiovascular (ISCV) Vulnerabilities: Improper Authentication, Use of Weak Credentials 2. RISK EVALUATION Successful exploitation of these vulnerabilities could allow an attacker to replay the session of the logged in ISCV user and gain access to patient records. 3. TECHNICAL DETAILS 3.1 AFFECTED PRODUCTS Philips reports the following versions of Intellispace Cardiovascular (ISC
     

Philips Intellispace Cardiovascular (ISCV)

Por:CISA
13 de Março de 2025, 09:52

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 8.5
  • ATTENTION: Low attack complexity
  • Vendor: Philips
  • Equipment: Intellispace Cardiovascular (ISCV)
  • Vulnerabilities: Improper Authentication, Use of Weak Credentials

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker to replay the session of the logged in ISCV user and gain access to patient records.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

Philips reports the following versions of Intellispace Cardiovascular (ISCV), an image and information management product, are affected:

  • Intellispace Cardiovascular (ISCV): Version 4.1 and prior (CVE-2025-2229)
  • Intellispace Cardiovascular (ISCV): Version 5.1 and prior (CVE-2025-2230)

3.2 VULNERABILITY OVERVIEW

3.2.1 Improper Authentication CWE-287

A flaw exists in the Windows login flow where an AuthContext token can be exploited for replay attacks and authentication bypass.

CVE-2025-2230 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.7 has been calculated; the CVSS vector string is (CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N).

A CVSS v4 score has also been calculated for CVE-2025-2230. A base score of 8.5 has been calculated; the CVSS vector string is (CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N).

3.2.2 Use of Weak Credentials CWE-1391

A token is created using the username, current date/time, and a fixed AES-128 encryption key, which is the same across all installations.

CVE-2025-2229 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.7 has been calculated; the CVSS vector string is (CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N).

A CVSS v4 score has also been calculated for CVE-2025-2229. A base score of 8.5 has been calculated; the CVSS vector string is (CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Netherlands

3.4 RESEARCHER

Joe Dillon reported these vulnerabilities to Philips.

4. MITIGATIONS

Philips recommends the following mitigations:

  • CVE-2025-2229: Resolved in ISCV 4.2 build 20589, which was released in May 2019.
  • CVE-2025-2230: Resolved in ISCV 5.2, which was released in September 2020.
  • Philips recommends users upgrade ISCV installed base to the latest ISCV version (at the time of this publication is 830089 – IntelliSpace Cardiovacular 8.0.0.0)
  • Please contact a local Philips sales (service) representative to learn how to engage this upgrade process.
  • For managed services users, new releases will be made available upon resource availability. Releases are subject to country-specific regulations.

Refer to the Philips advisory for more details.

CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the Internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time. These vulnerabilities are not exploitable remotely.

5. UPDATE HISTORY

  • March 13, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • Dario Health USB-C Blood Glucose Monitoring System Starter Kit Android Application CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 8.7 ATTENTION: Exploitable remotely/low attack complexity Vendor: Dario Health Equipment: USB-C Blood Glucose Monitoring System Starter Kit Android Application, Application Database and Internet-based Server Infrastructure Vulnerabilities: Exposure of Private Personal Information to an Unauthorized Actor, Improper Output Neutralization For Logs, Storage of Sensitive Data In a Mechanism Without Access Control, Cleartext Transmission of Sensitive Information
     

Dario Health USB-C Blood Glucose Monitoring System Starter Kit Android Application

Por:CISA
27 de Fevereiro de 2025, 09:57

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 8.7
  • ATTENTION: Exploitable remotely/low attack complexity
  • Vendor: Dario Health
  • Equipment: USB-C Blood Glucose Monitoring System Starter Kit Android Application, Application Database and Internet-based Server Infrastructure
  • Vulnerabilities: Exposure of Private Personal Information to an Unauthorized Actor, Improper Output Neutralization For Logs, Storage of Sensitive Data In a Mechanism Without Access Control, Cleartext Transmission of Sensitive Information, Cross-site Scripting (XSS), Sensitive Cookie Without 'HttpOnly' Flag, Exposure of Sensitive Information Due To Incompatible Policies

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker to expose information, inject code, manipulate data, or achieve cross-site scripting (XSS), resulting in full session compromise.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Dario Health products are affected:

  • USB-C Blood Glucose Monitoring System Starter Kit Android Applications: Versions 5.8.7.0.36 and prior
  • Dario Application Database and Internet-based Server Infrastructure: All versions

3.2 VULNERABILITY OVERVIEW

3.2.1 EXPOSURE OF PRIVATE PERSONAL INFORMATION TO AN UNAUTHORIZED ACTOR CWE-359

An attacker could expose cross-user Personal Identifiable Information (PII) and personal health information transmitted to the Android device via the Dario Health application database.

CVE-2025-20060 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.5 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N).

A CVSS v4 score has also been calculated for CVE-2025-20060. A base score of 8.7 has been calculated; the CVSS vector string is (CVSS4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N).

3.2.2 IMPROPER OUTPUT NEUTRALIZATION FOR LOGS CWE-117

Unauthenticated log effects metrics gathering incident response efforts and potentially exposes risk of injection attacks (ex log injection).

CVE-2025-23405 has been assigned to this vulnerability. A CVSS v3.1 base score of 5.3 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:L/A:N).

A CVSS v4 score has also been calculated for CVE-2025-23405. A base score of 6.9 has been calculated; the CVSS vector string is (CVSS4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:L/VA:N/SC:N/SI:L/SA:N).

3.2.3 STORAGE OF SENSITIVE DATA IN A MECHANISM WITHOUT ACCESS CONTROL CWE-921

Insecure file retrieval process that facilitates potential for file manipulation to affect product stability and confidentiality, integrity, authenticity, and attestation of stored data.

CVE-2025-24843 has been assigned to this vulnerability. A CVSS v3.1 base score of 5.1 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:N/UI:N/S:U/C:N/I:L/A:L).

A CVSS v4 score has also been calculated for CVE-2025-24843. A base score of 5.1 has been calculated; the CVSS vector string is (CVSS4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:N/VI:L/VA:L/SC:N/SI:N/SA:N).

3.2.4 CLEARTEXT TRANSMISSION OF SENSITIVE INFORMATION CWE-319

Lack of encryption in transit for cloud infrastructure facilitating potential for sensitive data manipulation or exposure.

CVE-2025-24849 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.1 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-24849. A base score of 7.5 has been calculated; the CVSS vector string is (CVSS4.0/AV:A/AC:H/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.2.5 IMPROPER NEUTRALIZATION OF INPUT DURING WEB PAGE GENERATION ('CROSS-SITE SCRIPTING') CWE-79

The Dario Health portal service application is vulnerable to XSS, which could allow an attacker to obtain sensitive information.

CVE-2025-20049 has been assigned to this vulnerability. A CVSS v3.1 base score of 5.8 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:L/UI:R/S:C/C:H/I:N/A:N).

A CVSS v4 score has also been calculated for CVE-2025-20049. A base score of 7.1 has been calculated; the CVSS vector string is (CVSS4.0/AV:N/AC:H/AT:P/PR:L/UI:A/VC:H/VI:N/VA:N/SC:H/SI:H/SA:H).

3.2.6 SENSITIVE COOKIE WITHOUT 'HTTPONLY' FLAG CWE-1004

Cookie policy is observable via built-in browser tools. In the presence of XSS, this could lead to full session compromise.

CVE-2025-24318 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.8 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:N/UI:N/S:C/C:H/I:N/A:N).

A CVSS v4 score has also been calculated for CVE-2025-24318. A base score of 5.9 has been calculated; the CVSS vector string is (CVSS4.0/AV:N/AC:H/AT:P/PR:N/UI:A/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N).

3.2.7 EXPOSURE OF SENSITIVE INFORMATION DUE TO INCOMPATIBLE POLICIES CWE-213

The Dario Health Internet-based server infrastructure is vulnerable due to exposure of development environment details, which could lead to unsafe functionality.

CVE-2025-24316 has been assigned to this vulnerability. A CVSS v3.1 base score of 5.3 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N).

A CVSS v4 score has also been calculated for CVE-2025-24316. A base score of 6.9 has been calculated; the CVSS vector string is (CVSS4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:N/VA:N/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: United States

3.4 RESEARCHER

Noah Cutler and Manuel Del Rio of Accenture reported these vulnerabilities to CISA.

4. MITIGATIONS

Dario Health recommends users update their Dario Health Android mobile application to the latest version. No other actions are required by users.

Dario Health recommends users perform the following mitigations:

  • Update the application from trusted sources.
  • Don't use rooted/jailbroken devices.
  • Avoid public untrusted networks

    For more information contact Dario Health directly.

CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the Internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time.

5. UPDATE HISTORY

  • February 27, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • Medixant RadiAnt DICOM Viewer CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 5.7 ATTENTION: Low attack complexity Vendor: Medixant Equipment: RadiAnt DICOM Viewer Vulnerability: Improper Certificate Validation 2. RISK EVALUATION Successful exploitation of this vulnerability could allow an attacker to perform a machine-in-the-middle attack (MITM), resulting in malicious updates being delivered to the user. 3. TECHNICAL DETAILS 3.1 AFFECTED PRODUCTS The following Medixant products are affected: RadiAnt DICOM Viewer: Version 2024.02
     

Medixant RadiAnt DICOM Viewer

Por:CISA
20 de Fevereiro de 2025, 10:03

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 5.7
  • ATTENTION: Low attack complexity
  • Vendor: Medixant
  • Equipment: RadiAnt DICOM Viewer
  • Vulnerability: Improper Certificate Validation

2. RISK EVALUATION

Successful exploitation of this vulnerability could allow an attacker to perform a machine-in-the-middle attack (MITM), resulting in malicious updates being delivered to the user.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Medixant products are affected:

  • RadiAnt DICOM Viewer: Version 2024.02

3.2 VULNERABILITY OVERVIEW

3.2.1 IMPROPER CERTIFICATE VALIDATION CWE-295

The affected product is vulnerable due to failure of the update mechanism to verify the update server's certificate which could allow an attacker to alter network traffic and carry out a machine-in-the-middle attack (MITM). An attacker could modify the server's response and deliver a malicious update to the user.

CVE-2025-1001 has been assigned to this vulnerability. A CVSS v3.1 base score of 5.7 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:H/A:N).

A CVSS v4 score has also been calculated for CVE-2025-1001. A base score of 5.7 has been calculated; the CVSS vector string is (AV:A/AC:L/AT:P/PR:N/UI:A/VC:N/VI:H/VA:N/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Poland

3.4 RESEARCHER

Sharon Brizinov of Claroty Team82 reported this vulnerability to CISA.

4. MITIGATIONS

Medixant recommends users download the v2025.1 or later version of their software.

If users are unable to update to the new version, Medixant recommends the following:

  • Disable the display of available updates via this command reg add "HKCU\Software\RadiAnt Viewer" /t REG_DWORD /v CheckUpdate /d 0 /f.
  • Do not check manually for updates ("Check for updates now" from the toolbar menu).
  • Ignore any update notifications coming from RadiAnt DICOM Viewer, download the latest version directly in the web browser from https://www.radiantviewer.com.
  • Check the downloaded RadiAnt DICOM Viewer installation package with antivirus software before running it.

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time. This vulnerability is not exploitable remotely.

5. UPDATE HISTORY

  • February 20, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • Qardio Heart Health IOS and Android Application and QardioARM A100 CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 7.2 ATTENTION: Low attack complexity Vendor: Qardio Equipment: Heart Health IOS application, Heart Health Android Application, QardioARM A100 Vulnerabilities: Exposure of Private Personal Information to an Unauthorized Actor, Uncaught Exception, Files or Directories Accessible to External Parties 2. RISK EVALUATION Successful exploitation of these vulnerabilities could allow an attacker to obtain sensitive information, cause a denial-of-service condition,
     

Qardio Heart Health IOS and Android Application and QardioARM A100

Por:CISA
13 de Fevereiro de 2025, 10:46

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 7.2
  • ATTENTION: Low attack complexity
  • Vendor: Qardio
  • Equipment: Heart Health IOS application, Heart Health Android Application, QardioARM A100
  • Vulnerabilities: Exposure of Private Personal Information to an Unauthorized Actor, Uncaught Exception, Files or Directories Accessible to External Parties

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker to obtain sensitive information, cause a denial-of-service condition, and obtain firmware files.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Qardio products are affected:

  • Qardio Heart Health IOS Mobile Application: Version 2.7.4
  • Qardio Heart Health Android Mobile Application: Version 2.5.1
  • QardioARM A100: All versions

3.2 VULNERABILITY OVEERVIEW

3.2.1 EXPOSURE OF PRIVATE PERSONAL INFORMATION TO AN UNAUTHORIZED ACTOR CWE-359

The Qardio Arm iOS application exposes sensitive data such as usernames and passwords in a plist file. This allows an attacker to log in to production-level development accounts and access an engineering backdoor in the application. The engineering backdoor allows the attacker to send hex-based commands over a UI-based terminal.

CVE-2025-20615 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.2 has been calculated; the CVSS vector string is (AV:P/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:L).

A CVSS v4 score has also been calculated for CVE-2025-20615. A base score of 6.9 has been calculated; the CVSS vector string is (AV:P/AC:L/AT:N/PR:L/UI:N/VC:H/VI:H/VA:L/SC:N/SI:N/SA:N).

3.2.2 UNCAUGHT EXCEPTION CWE-248

With a specially crafted Python script, an attacker could send continuous startMeasurement commands over an unencrypted Bluetooth connection to the affected device. This would prevent the device from connecting to a clinician's app to take patient readings and ostensibly flood it with requests, resulting in a denial-of-service condition.

CVE-2025-24836 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.1 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-24836. A base score of 7.2 has been calculated; the CVSS vector string is (AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:H/VA:H/SC:N/SI:N/SA:N).

3.2.3 FILES OR DIRECTORIES ACCESSIBLE TO EXTERNAL PARTIES CWE-552

An attacker could obtain firmware files and reverse engineer their intended use leading to loss of confidentiality and integrity of the hardware devices enabled by the Qardio iOS and Android applications.

CVE-2025-23421 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.4 has been calculated; the CVSS vector string is (AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:L).

A CVSS v4 score has also been calculated for CVE-2025-23421. A base score of 6.9 has been calculated; the CVSS vector string is (AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:L/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: United States

3.4 RESEARCHER

Bryan Riggins of Insulet Corporation reported these vulnerabilities to CISA.

4. MITIGATIONS

Qardio has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Qardio customer support for additional information.

Users should do the following to help mitigate the risk:

  • Disable Bluetooth when not in use.
  • Don't use this device in public or within Bluetooth range of malicious actors.
  • Only use trusted mobile apps from trusted providers.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time. These vulnerabilities is are not exploitable remotely.

5. UPDATE HISTORY

  • February 13, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • Orthanc Server CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 9.2 ATTENTION: Exploitable remotely/low attack complexity Vendor: Orthanc Equipment: Orthanc Server Vulnerability: Missing Authentication for Critical Function 2. RISK EVALUATION Successful exploitation of this vulnerability could allow an attacker to disclose sensitive information, modify records, or cause a denial-of-service condition. 3. TECHNICAL DETAILS 3.1 AFFECTED PRODUCTS The following Orthanc products are affected: Orthanc server: Versions prior
     

Orthanc Server

Por:CISA
6 de Fevereiro de 2025, 10:19

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 9.2
  • ATTENTION: Exploitable remotely/low attack complexity
  • Vendor: Orthanc
  • Equipment: Orthanc Server
  • Vulnerability: Missing Authentication for Critical Function

2. RISK EVALUATION

Successful exploitation of this vulnerability could allow an attacker to disclose sensitive information, modify records, or cause a denial-of-service condition.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Orthanc products are affected:

  • Orthanc server: Versions prior to 1.5.8

3.2 VULNERABILITY OVERVIEW

3.2.1 MISSING AUTHENTICATION FOR CRITICAL FUNCTION CWE-306

Orthanc server prior to version 1.5.8 does not enable basic authentication by default when remote access is enabled. This could result in unauthorized access by an attacker.

CVE-2025-0896 has been assigned to this vulnerability. A CVSS v3.1 base score of 9.8 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-0896. A base score of 9.2 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Belgium

3.4 RESEARCHER

Amitay Dan reported this vulnerability to Orthanc.
Souvik Kandar reported this vulnerability to CISA.

4. MITIGATIONS

Orthanc recommends that users update to the latest version or enable the HTTP authentication by setting the configuration "AuthenticationEnabled": true in the configuration file.

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, utilize more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have their own vulnerabilities and should be updated to the most current version available. Also recognize VPNs are only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time.

5. UPDATE HISTORY

  • February 6, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • MicroDicom DICOM Viewer CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 5.7 ATTENTION: Low attack complexity Vendor: MicroDicom Equipment: DICOM Viewer Vulnerability: Improper Certificate Validation 2. RISK EVALUATION Successful exploitation of this vulnerability could allow an attacker to alter network traffic and perform a machine-in-the-middle (MITM) attack. 3. TECHNICAL DETAILS 3.1 AFFECTED PRODUCTS The following MicroDicom DICOM Viewer are affected: MicroDicom DICOM Viewer: Version 2024.03 3.2 VULNERABILITY OVERVIEW 3.
     

MicroDicom DICOM Viewer

Por:CISA
6 de Fevereiro de 2025, 10:13

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 5.7
  • ATTENTION: Low attack complexity
  • Vendor: MicroDicom
  • Equipment: DICOM Viewer
  • Vulnerability: Improper Certificate Validation

2. RISK EVALUATION

Successful exploitation of this vulnerability could allow an attacker to alter network traffic and perform a machine-in-the-middle (MITM) attack.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following MicroDicom DICOM Viewer are affected:

  • MicroDicom DICOM Viewer: Version 2024.03

3.2 VULNERABILITY OVERVIEW

3.2.1 IMPROPER CERTIFICATE VALIDATION CWE-295

MicroDicom DICOM Viewer fails to adequately verify the update server's certificate, which could make it possible for attackers in a privileged network position to alter network traffic and carry out a machine-in-the-middle (MITM) attack. This allows the attackers to modify the server's response and deliver a malicious update to the user.

CVE-2025-1002 has been assigned to this vulnerability. A CVSS v3.1 base score of 5.7 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:H/A:N).

A CVSS v4 score has also been calculated for CVE-2025-1002. A base score of 5.7 has been calculated; the CVSS vector string is (AV:A/AC:L/AT:P/PR:N/UI:A/VC:N/VI:H/VA:N/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Bulgaria

3.4 RESEARCHER

Sharon Brizinov of Claroty Team82 reported this vulnerability to CISA.

4. MITIGATIONS

MicroDicom recommends users upgrade to DICOM Viewer version 2025.1.

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time. This vulnerability is not exploitable remotely.

5. UPDATE HISTORY

  • February 6, 2025: Initial Publication
  • ✇ICS Medical Advisories
  • Contec Health CMS8000 Patient Monitor (Update A) CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 9.3 ATTENTION: Exploitable remotely/low attack complexity Vendor: Contec Health Equipment: CMS8000 Patient Monitor Vulnerabilities: Out-of-Bounds Write, Hidden Functionality, Privacy Leakage 2. RISK EVALUATION Successful exploitation of these vulnerabilities could allow an attacker to remotely send specially formatted UDP requests or connect to an unknown external network that would allow them to write arbitrary data, resulting in remote code execution. T
     

Contec Health CMS8000 Patient Monitor (Update A)

Por:CISA
30 de Janeiro de 2025, 14:08

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 9.3
  • ATTENTION: Exploitable remotely/low attack complexity
  • Vendor: Contec Health
  • Equipment: CMS8000 Patient Monitor
  • Vulnerabilities: Out-of-Bounds Write, Hidden Functionality, Privacy Leakage

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker to remotely send specially formatted UDP requests or connect to an unknown external network that would allow them to write arbitrary data, resulting in remote code execution. The device may also leak patient information and sensor data to the same unknown external network. Simultaneous exploitation of all vulnerable devices on a shared network is possible.

The Food and Drug Administration (FDA) has released a safety communication in connection with these vulnerabilities.

CISA has released an additional Fact Sheet for CVE-2025-0626 and CVE-2025-0683.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Contec Health products are affected:

  • CMS8000 Patient Monitor Firmware: Version smart3250-2.6.27-wlan2.1.7.cramfs
  • CMS8000 Patient Monitor Firmware: Version CMS7.820.075.08/0.74(0.75)
  • CMS8000 Patient Monitor Firmware: Version CMS7.820.120.01/0.93(0.95)
  • CMS8000 Patient Monitor: All versions (CVE-2025-0626, CVE-2025-0683, CVE-2025-1204)

3.2 VULNERABILITY OVERVIEW

3.2.1 OUT-OF-BOUNDS WRITE CWE-787

The affected product is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution.

CVE-2024-12248 has been assigned to this vulnerability. A CVSS v3.1 base score of 9.8 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2024-12248. A base score of 9.3 has been calculated; the CVSS vector string is (CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.2.2 HIDDEN FUNCTIONALITY (BACKDOOR) CWE-912

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

CVE-2025-0626 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.5 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-0626. A base score of 7.7 has been calculated; the CVSS vector string is (CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.2.3 EXPOSURE OF PRIVATE PERSONAL INFORMATION TO AN UNAUTHORIZED ACTOR (PRIVACY LEAKAGE) CWE-359

In its default configuration, the affected product transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

CVE-2025-0683 has been assigned to this vulnerability. A CVSS v3.1 base score of 5.9 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N).

A CVSS v4 score has also been calculated for CVE-2025-0683. A base score of 8.2 has been calculated; the CVSS vector string is (CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N).

3.2.4 HIDDEN FUNCTIONALITY CWE-912

The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.

CVE-2025-1204 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.5 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-1204. A base score of 7.7 has been calculated; the CVSS vector string is (CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: China

3.4 RESEARCHER

An anonymous researcher reported CVE-2024-12248, CVE-2025-0626, and CVE-2025-0683 to CISA.

Claroty - Team82 reported CVE-2025-1024 to CISA.

4. MITIGATIONS

Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.

If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.

Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication.

CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks. Update firewall rules to prevent access to potentially affected devices.
  • If network connected, ensure all medical devices are on a separate, low privilege subnet.
  • Only use trusted manufacturers for safety critical systems.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time.

5. UPDATE HISTORY

  • January 30, 2025: Initial Publication
  • February 25, 2025: Update A - Added CVE-2025-1204, updated description for CVE-2025-0626. Added mitigations for the specific IP addresses. Added researcher credit to Claroty for CVE-2025-1204.
  • ✇ICS Medical Advisories
  • Ossur Mobile Logic Application CISA
    View CSAF 1. EXECUTIVE SUMMARY CVSS v4 5.6 ATTENTION: Low attack complexity Vendor: Ossur Equipment: Mobile Logic Application Vulnerabilities: Exposure of Sensitive System Information to an Unauthorized Control Sphere, Command Injection, Use of Hard-coded Credentials 2. RISK EVALUATION Successful exploitation of these vulnerabilities could allow an attacker unauthorized access to sensitive information. 3. TECHNICAL DETAILS 3.1 AFFECTED PRODUCTS The following Ossur products are affected: Logic
     

Ossur Mobile Logic Application

Por:CISA
19 de Dezembro de 2024, 10:56

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 5.6
  • ATTENTION: Low attack complexity
  • Vendor: Ossur
  • Equipment: Mobile Logic Application
  • Vulnerabilities: Exposure of Sensitive System Information to an Unauthorized Control Sphere, Command Injection, Use of Hard-coded Credentials

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker unauthorized access to sensitive information.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Ossur products are affected:

  • Logic Mobile Application: Versions prior to 1.5.5

3.2 Vulnerability Overview

3.2.1 EXPOSURE OF SENSITIVE SYSTEM INFORMATION TO AN UNAUTHORIZED CONTROL SPHERE CWE-497

A valid set of credentials in a .js file and a static token for communication were obtained from the decompiled IPA. An attacker could use the information to disrupt normal use of the application by changing the translation files and thus weaken the integrity of normal use.

CVE-2024-53683 has been assigned to this vulnerability. A CVSS v3.1 base score of 4.4 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N).

A CVSS v4 score has also been calculated for CVE-2024-53683. A base score of 5.6 has been calculated; the CVSS vector string is (CVSS4.0/AV:L/AC:L/AT:P/PR:H/UI:N/VC:N/VI:H/VA:N/SC:N/SI:N/SA:N).

3.2.2 IMPROPER NEUTRALIZATION OF SPECIAL ELEMENTS USED IN A COMMAND ('COMMAND INJECTION') CWE-77

Multiple bash files were present in the application's private directory. Bash files can be used on their own, by an attacker that has already full access to the mobile platform to compromise the translations for the application.

CVE-2024-54681 has been assigned to this vulnerability. A CVSS v3.1 base score of 3.5 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:L/UI:R/S:U/C:N/I:N/A:L).

A CVSS v4 score has also been calculated for CVE-2024-54681. A base score of 2.0 has been calculated; the CVSS vector string is (CVSS4.0/AV:N/AC:L/AT:P/PR:L/UI:A/VC:N/VI:N/VA:L/SC:N/SI:N/SA:N).

3.2.3 USE OF HARD-CODED CREDENTIALS CWE-798

Hard-coded credentials were included as part of the application binary. These credentials served as part of the application authentication flow and communication with the mobile application. An attacker could access unauthorized information.

CVE-2024-45832 has been assigned to this vulnerability. A CVSS v3.1 base score of 4.3 has been calculated; the CVSS vector string is (AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L).

A CVSS v4 score has also been calculated for CVE-2024-45832. A base score of 2.0 has been calculated; the CVSS vector string is (CVSS4.0/AV:P/AC:L/AT:P/PR:N/UI:N/VC:L/VI:L/VA:L/SC:L/SI:H/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Iceland

3.4 RESEARCHER

Bryan Riggins reported these vulnerabilities to CISA.

4. MITIGATIONS

Ossur recommends users download Version 1.5.5 or later of the mobile application. The latest version of the application can be obtained through the app store on respective mobile devices. No additional action is required by users.

CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the Internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time. These vulnerabilities are not exploitable remotely.

5. UPDATE HISTORY

  • December 19, 2024: Initial Publication
❌
❌